Abstract
Electronic portal imaging devices (EPIDs) have been the preferred tools for verification
of patient positioning for radiotherapy in recent decades. Since EPID images contain
dose information, many groups have investigated their use for radiotherapy dose measurement.
With the introduction of the amorphous-silicon EPIDs, the interest in EPID dosimetry
has been accelerated because of the favourable characteristics such as fast image
acquisition, high resolution, digital format, and potential for in vivo measurements
and 3D dose verification. As a result, the number of publications dealing with EPID
dosimetry has increased considerably over the past ∼15 years. The purpose of this
paper was to review the information provided in these publications. Information available
in the literature included dosimetric characteristics and calibration procedures of
various types of EPIDs, strategies to use EPIDs for dose verification, clinical approaches
to EPID dosimetry, ranging from point dose to full 3D dose distribution verification,
and current clinical experience. Quality control of a linear accelerator, pre-treatment
dose verification and in vivo dosimetry using EPIDs are now routinely used in a growing
number of clinics. The use of EPIDs for dosimetry purposes has matured and is now
a reliable and accurate dose verification method that can be used in a large number
of situations. Methods to integrate 3D in vivo dosimetry and image-guided radiotherapy
(IGRT) procedures, such as the use of kV or MV cone-beam CT, are under development.
It has been shown that EPID dosimetry can play an integral role in the total chain
of verification procedures that are implemented in a radiotherapy department. It provides
a safety net for simple to advanced treatments, as well as a full account of the dose
delivered. Despite these favourable characteristics and the vast range of publications
on the subject, there is still a lack of commercially available solutions for EPID
dosimetry. As strategies evolve and commercial products become available, EPID dosimetry
has the potential to become an accurate and efficient means of large-scale patient-specific
IMRT dose verification for any radiotherapy department.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Radiotherapy and OncologyAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Clinical use of electronic portal imaging.Semin Radiat Oncol. 2005; 15: 157-167
- Clinical use of electronic portal imaging: report of AAPM Radiation Therapy Committee Task Group 58.Med Phys. 2001; 28: 712-737
- Guide to clinical use of electronic portal imaging.J Appl Clin Med Phys. 2000; 1: 38-57
- Portal imaging.Br J Radiol. 2001; 74: 789-804
- Electronic portal imaging devices: a review and historical perspective of contemporary technologies and research.Phys Med Biol. 2002; 47: R31-65
- QA-QC of IMRT: European perspective.in: Bortfeld T. Schmidt-Ullrich R. De Neve W. Wazer D.E. Image-guided IMRT. Springer-Verlag, Berlin-Heidelberg2006
- QA-QC of IMRT: American perspective.in: Bortfeld T. Schmidt-Ullrich R. De Neve W. Wazer D.E. Image-guided IMRT. Springer-Verlag, Berlin-Heidelberg2006
- Dosimetric quality assurance for intensity-modulated radiotherapy feasibility study for a filmless approach.Strahlenther Onkol. 2005; 181: 468-474
- Dosimetric characterization of a large area pixel-segmented ionization chamber.Med Phys. 2004; 31: 414-420
- Characterization of a 2D ion chamber array for the verification of radiotherapy treatments.Phys Med Biol. 2005; 50: 3361-3373
- Evaluation of a 2D diode array for IMRT quality assurance.Radiother Oncol. 2004; 70: 199-206
- A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.Med Phys. 2003; 30: 870-879
- Dosimetric verification of 6 and 18 MV intensity modulated photon beams using a dedicated fluoroscopic electronic portal imaging device (EPID).Radiother Oncol. 2006; 81: 88-96
- Tomotherapeutic portal imaging for radiation treatment verification.Phys Med Biol. 1998; 43: 3607-3616
- Performance of electronic portal imaging devices (EPIDs) used in radiotherapy: image quality and dose measurements.Med Phys. 2004; 31: 985-996
- A review of electronic portal imaging devices (EPIDs).Med Phys. 1992; 19: 1-16
- Developments in electronic portal imaging systems.Br J Radiol. 2006; 79 Spec No. 1: S50-S65
- Physical aspects of a liquid-filled ionization chamber with pulsed polarizing voltage.Med Phys. 1991; 18: 692-702
- First clinical experience with a newly developed electronic portal imaging device.Int J Rad Oncol Biol Phys. 1990; 18: 1173-1181
- A liquid ionisation detector for digital radiography of therapeutic megavoltage photon beams.Phys Med Biol. 1985; 30: 313-321
- A matrix ionisation chamber imaging device for on-line patient setup verification during radiotherapy.Radiother Oncol. 1988; 11: 369-378
- Transmission dosimetry with a liquid-filled electronic portal imaging device.Int J Rad Oncol Biol Phys. 1996; 34: 931-941
- Input/output characteristics of a matrix ion-chamber electronic portal imaging device.Med Phys. 1994; 21: 1447-1454
- Dosimetric characteristics of a liquid-filled electronic portal imaging device.Int J Rad Oncol Biol Physics. 1995; 33: 1265-1272
- Portal dosimetry using a liquid ion chamber matrix: dose response studies.Med Phys. 1995; 22: 1101-1106
- The dose response relationship of a liquid-filled electronic portal imaging device.Med Phys. 1996; 23: 1601-1611
- Calibration of a portal imaging device for high-precision dosimetry: a Monte Carlo study.Med Phys. 1998; 25: 1891-1902
- The use of an electronic portal imaging system to measure portal dose and portal dose profiles.Med Phys. 1998; 25: 1903-1909
- An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factor.Med Phys. 2001; 28: 2247-2257
- Pre-treatment dosimetric verification by means of a liquid-filled electronic portal imaging device during dynamic delivery of intensity modulated treatment fields.Radiother Oncol. 2001; 60: 181-190
- Evaluation of rapid dose map acquisition of a scanning liquid-filled ionization chamber electronic portal imaging device.Int J Rad Oncol Biol Phys. 2003; 55: 1432-1445
- Correction for dose–response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle.Phys Med Biol. 2004; 49: N93-103
- The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes.Med Phys. 2004; 31: 819-827
- The prediction of transmitted dose distributions using a 3D treatment planning system.Australas Phys Eng Sci Med. 2006; 29: 18-29
- Two-dimensional transmitted dose measurements using a scanning liquid ionization chamber EPID.Phys Med Biol. 2006; 51: 2971-2985
- Dosimetric properties of the liquid ionization chamber electronic portal imaging device (EPID).Igaku Butsuri. 2006; 26: 28-38
- A linear array, scintillation crystal-photodiode detector for megavoltage imaging.Med Phys. 1991; 18: 681-691
- An electronic portal imaging device for transit dosimetry.Phys Med Biol. 1997; 42: 2273-2283
- The application of transit dosimetry to precision radiotherapy.Med Phys. 1996; 23: 713-721
- Rapid portal imaging with a high-efficiency, large field-of-view detector.Med Phys. 1998; 25: 2333-2346
- Physical characteristics of a commercial electronic portal imaging device.Med Phys. 1996; 23: 1845-1855
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID).Phys Med Biol. 1995; 40: 1943-1955
- Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation.Phys Med Biol. 1998; 43: 2047-2060
- Verification of treatment parameter transfer by means of electronic portal dosimetry.Med Phys. 2004; 31: 341-347
- Dosimetric properties of the Theraview fluoroscopic electronic portal imaging device.Br J Radiol. 2000; 73: 517-530
- Measurement of the response of Gd2O2S:Tb phosphor to 6 MV X-rays.Phys Med Biol. 2001; 46: 517-530
- Characteristics relevant to portal dosimetry of a cooled CCD camera-based EPID.Med Phys. 2004; 31: 2549-2551
- A novel approach to accurate portal dosimetry using CCD-camera based EPIDs.Med Phys. 2006; 33: 888-903
- Characterization of a high-elbow, fluoroscopic electronic portal imaging device for portal dosimetry.Phys Med Biol. 2000; 45: 197-216
- Optical scattering in camera-based electronic portal imaging.Phys Med Biol. 1999; 44: 2381-2396
- IMRT verification with a camera-based electronic portal imaging system.Phys Med Biol. 2000; 45: N183-N196
- Testing of the stability of intensity modulated beams generated with dynamic multileaf collimation, applied to the MM50 racetrack microtron.Med Phys. 2000; 27: 2701-2707
- Daily dosimetric quality control of the MM50 racetrack microtron using an electronic portal imaging device.Radiother Oncol. 1995; 37: 55-60
- Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging.Int J Radiat Oncol Biol Phys. 1998; 42: 437-454
- A real-time, flat-panel, amorphous silicon, digital X-ray imager.Radiographics. 1995; 15: 993-1000
- Dose–response and ghosting effects of an amorphous silicon electronic portal imaging device.Med Phys. 2004; 31: 285-295
- An intercomparison of 11 amorphous silicon EPIDs of the same type: implications for portal dosimetry.Phys Med Biol. 2006; 51: 4189-4200
- Monte Carlo modelling of a-Si EPID response: the effect of spectral variations with field size and position.Med Phys. 2006; 33: 4527-4540
- Dose–response characteristics of an amorphous silicon EPID.Med Phys. 2005; 32: 3095-3105
- Amorphous silicon EPID calibration for dosimetric applications: comparison of a method based on Monte Carlo prediction of response with existing techniques.Phys Med Biol. 2007; 52: 3351-3368
- Calibration of an amorphous-silicon flat panel portal imager for exit-beam dosimetry.Med Phys. 2006; 33: 584-594
- A global calibration model for a-Si EPIDs used for transit dosimetry.Med Phys. 2007; 34: 3872-3884
- Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry.Med Phys. 2005; 32: 3558-3568
- The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields.Radiother Oncol. 2004; 71: 223-234
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.Med Phys. 2003; 30: 1618-1627
- Consequences of the spectral response of an a-Si EPID and implications for dosimetric calibration.Med Phys. 2005; 32: 2649-2658
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device.Med Phys. 2001; 28: 911-924
- Monte Carlo computation of dosimetric amorphous silicon electronic portal images.Med Phys. 2004; 31: 2135-2146
- An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry.Med Phys. 2002; 29: 2262-2268
- Comprehensive Monte Carlo calculation of the point spread function for a commercial a-Si EPID.Med Phys. 2005; 32: 1115-1127
- The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes.Med Phys. 2004; 31: 2989-2995
- Relative dosimetry using active matrix flat-panel imager (AMFPI) technology.Med Phys. 1999; 26: 1530-1541
- A performance comparison of direct- and indirect-detection flat-panel imagers.Nucl Instrum Methods Phys Res A. 2002; 484: 351-363
- Blurring artifacts in megavoltage radiography with a flat-panel imaging system: comparison of Monte Carlo simulations with measurements.Med Phys. 2002; 29: 2559-2570
- Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device.Med Phys. 2007; 34: 3815-3824
- Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method.Med Phys. 2006; 33: 259-273
- Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID.Med Phys. 2006; 33: 4367-4378
- Comparison of ghosting effects for three commercial a-Si EPIDs.Med Phys. 2006; 33: 2448-2451
- A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager.Med Phys. 1999; 26: 1624-1641
- Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager.Med Phys. 2005; 32: 3267-3278
- A convolution model to convert transmission dose images to exit dose distributions.Med Phys. 1997; 24: 189-199
- Relative profile and dose verification of intensity-modulated radiation therapy.Int J Rad Oncol Biol Phys. 2000; 47: 231-240
- Monte Carlo simulation of portal dosimetry on a rectilinear voxel geometry: a variable gantry angle solution.Phys Med Biol. 2003; 48: N231-238
- Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.Phys Med Biol. 2005; 50: 1377-1392
- An electronic portal imaging device as a physics tool.Med Dosim. 1997; 22: 101-105
- Fast, daily linac verification for segmented IMRT using electronic portal imaging.Radiother Oncol. 2006; 80: 86-92
- Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID.Phys Med Biol. 2007; 52: 1721-1733
- Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment.Med Phys. 2004; 31: 2091-2096
- Assessment of flatness and symmetry of megavoltage X-ray beam with an electronic portal imaging device (EPID).Australas Phys Eng Sci Med. 2002; 25: 58-66
- Intensity-modulated radiation therapy: overlapping co-axial modulated fields.Phys Med Biol. 2004; 49: 3629-3637
- Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning system.Phys Med Biol. 2006; 51: N65-78
- Accurate measurement of the dynamic response of a scanning electronic portal imaging device.Med Phys. 2001; 28: 310-316
- Utilizing an electronic portal imaging device to monitor light and radiation field congruence.J Appl Clin Med Phys. 2003; 4: 315-320
- Computer verification of fluence map for intensity modulated radiation therapy.Med Phys. 2000; 27: 2084-2092
- Quantitative measurement of MLC leaf displacements using an electronic portal image device.Phys Med Biol. 2004; 49: 1521-1533
- Verification of multileaf collimator leaf positions using an electronic portal imaging device.Med Phys. 2002; 29: 2900-2912
- Evaluation of two methods of predicting MLC leaf positions using EPID measurements.Med Phys. 2006; 33: 3174-3182
- Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging device.Med Phys. 2002; 29: 2034-2040
- Leaf position verification during dynamic beam delivery: a comparison of three applications using electronic portal imaging.Med Phys. 2000; 27: 1601-1609
- A method for geometrical verification of dynamic intensity modulated radiotherapy using a scanning electronic portal imaging device.Med Phys. 2002; 29: 1071-1079
- Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager.Med Phys. 2004; 31: 389-395
- Verification of dynamic multileaf collimation using an electronic portal imaging device.Phys Med Biol. 2000; 45: 495-509
- Using fluence separation to account for energy spectra dependence in computing dosimetric a-Si EPID images for IMRT fields.Med Phys. 2006; 33: 4468-4480
- Investigation of tilted dose kernels for portal dose prediction in a-Si electronic portal imagers.Med Phys. 2006; 33: 3333-3339
- Monte Carlo studies of the exit photon spectra and dose to a metal/phosphor portal imaging screen.Med Phys. 2000; 27: 330-339
- Using the frame averaging of aS500 EPID for IMRT verification.J Appl Clin Med Phys. 2003; 4: 287-299
- On the verification of the incident energy fluence in tomotherapy IMRT.Phys Med Biol. 2001; 46: 2953-2965
- Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device.Med Phys. 2004; 31: 463-476
- Dosimetric IMRT verification with a flat-panel EPID.Med Phys. 2003; 30: 3143-3155
- Pre-treatment verification of intensity modulated photon beams with films and electronic portal imaging–two years of clinical experience.Z Med Phys. 2004; 14: 239-250
- An active matrix flat panel dosimeter (AMFPD) for in-phantom dosimetric measurements.Med Phys. 2005; 32: 466-472
- GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications.Med Phys. 2006; 33: 2839-2851
- 3D in vivo dosimetry by means of entrance dosimetry for dose guided radiotherapy.Radiother Oncol. 2006; 81: S46
- DAVID–a translucent multi-wire transmission ionization chamber for in vivo verification of IMRT and conformal irradiation techniques.Phys Med Biol. 2006; 51: 1237-1248
- Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans.Med Phys. 2006; 33: 3369-3382
- A Monte Carlo based three-dimensional dose reconstruction method derived from portal dose images.Med Phys. 2006; 33: 2426-2434
- Treatment verification in the presence of inhomogeneities using EPID-based three-dimensional dose reconstruction.Med Phys. 2007; 34: 2816-2826
- The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations.Radiother Oncol. 2008; 86: 86-92
- A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction.J Appl Clin Med Phys. 2005; 6: 22-39
- Three-dimensional IMRT verification with a flat-panel EPID.Med Phys. 2005; 32: 600-612
- 3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID.Radiother Oncol. 2007; 82: 201-207
- Correction of conebeam CT values using a planning CT for derivation of the “dose of the day”.Radiother Oncol. 2007; 85: 195-200
- Modeling dose distributions from portal dose images using the convolution/superposition method.Med Phys. 1996; 23: 1381-1392
- Analysis and convergence of the iterative convolution/superposition dose reconstruction technique for multiple treatment beams and tomotherapy.Med Phys. 1997; 24: 1465-1476
- Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure.Radiother Oncol. 1990; 17: 141-151
- The accuracy of CT-based inhomogeneity corrections and in vivo dosimetry for the treatment of lung cancer.Radiother Oncol. 1995; 37: 199-208
- Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients.Int J Rad Oncol Biol Phys. 1999; 45: 1297-1303
- Routine individualised patient dosimetry using electronic portal imaging devices.Radiother Oncol. 2007; 83: 65-75
- Primary photon fluence extraction from portal images acquired with an amorphous silicon flat panel detector: experimental determination of a scatter filter.J Exp Clin Cancer Res. 2007; 26: 125-132
- In vivo dosimetry by an aSi-based EPID.Med Phys. 2006; 33: 4414-4422
- Application of a practical method for the isocenter point in vivo dosimetry by a transit signal.Phys Med Biol. 2007; 52: 5101-5117
- Portal dose image (PDI) prediction for dosimetric treatment verification in radiotherapy. I. An algorithm for open beams.Med Phys. 1998; 25: 830-840
- Evaluation of an electronic portal imaging device for transit dosimetry.Acta Oncol. 1999; 38: 591-596
- Improvement of radiotherapy treatment delivery accuracy using an electronic portal imaging device.Radiat Prot Dosimetry. 2006; 121: 70-79
- Experimental verification of a portal dose prediction model.Med Phys. 2005; 32: 2805-2818
- Photon scatter in portal images: accuracy of a fluence based pencil beam superposition algorithm.Med Phys. 2000; 27: 913-922
- Photon scatter in portal images: physical characteristics of pencil beam kernels generated using the EGS Monte Carlo code.Med Phys. 2000; 27: 312-320
- A two-step algorithm for predicting portal dose images in arbitrary detectors.Med Phys. 2000; 27: 2109-2116
- Portal dose image verification: formalism and application of the collapsed cone superposition method.Phys Med Biol. 2002; 47: 4371-4387
- Portal dose image verification: the collapsed cone superposition method applied with different electronic portal imaging devices.Phys Med Biol. 2006; 51: 335-349
- Portal dose image prediction for dosimetric treatment verification in radiotherapy. II. An algorithm for wedged beams.Med Phys. 2002; 29: 925-931
- In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion.Radiother Oncol. 1998; 49: 125-132
- SIFT: a method to verify the IMRT fluence delivered during patient treatment using an electronic portal imaging device.Int J Rad Oncol Biol Phys. 2004; 60: 981-993
- Verification of patient position and delivery of IMRT by electronic portal imaging.Radiother Oncol. 2004; 73: 339-347
- Full forward Monte Carlo calculation of portal dose from MLC collimated treatment beams.Phys Med Biol. 2002; 47: 377-390
- Direct measurement and analytical modeling of scatter in portal imaging.Med Phys. 2000; 27: 462-471
- Analytical scatter kernels for portal imaging at 6 MV.Med Phys. 2001; 28: 553-559
- An iterative algorithm for reconstructing incident beam distributions from transmission measurements using electronic portal imaging.Phys Med Biol. 2001; 46: N203-211
- Scattered radiation in portal images: a Monte Carlo simulation and a simple physical model.Med Phys. 1996; 23: 63-73
- Extraction of primary signal from EPIDs using only forward convolution.Med Phys. 1997; 24: 1477-1484
- The practical implementation of a scatter model for portal imaging at 10 MV.Phys Med Biol. 1998; 43: 2685-2693
- Calculation of portal dose using the convolution/superposition method.Med Phys. 1996; 23: 527-535
- The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.J Appl Clin Med Phys. 2007; 8: 69-84
- Comparison of two-dimensional transmitted dose maps: evaluation of existing algorithms.Australas Phys Eng Sci Med. 2006; 29: 179-187
- Delivery verification in sequential and helical tomotherapy.Phys Med Biol. 1999; 44: 1815-1841
- The use of an electronic portal imaging device for exit dosimetry and quality control measurements.Int J Radiat Oncol Biol Phys. 1995; 31: 593-603
- Measurement possibilities using an electronic portal imaging device.Radiother Oncol. 1993; 29: 237-243
- Two-dimensional exit dosimetry using a liquid-filled electronic portal imaging device and a convolution model.Radiother Oncol. 1997; 44: 149-157
- New method to obtain the midplane dose using portal in vivo dosimetry.Int J Radiat Oncol Biol Phys. 1998; 41: 465-474
- First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose.Radiother Oncol. 1998; 47: 303-312
- Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.Med Phys. 2006; 33: 3921-3930
- Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT.Int J Radiat Oncol Biol Phys. 2007; 67: 1568-1577
- Patient-specific dosimetry of conventional and intensity modulated radiation therapy using a novel full Monte Carlo phase space reconstruction method from electronic portal images.Phys Med Biol. 2007; 52: 2277-2299
- IMRT verification by three-dimensional dose reconstruction from portal beam measurements.Med Phys. 2002; 29: 1847-1858
- Dose-guided radiation therapy with megavoltage cone-beam CT.Br J Radiol. 2006; 79 Spec No. 1: S87-98
- Three-dimensional dose reconstruction of breast cancer treatment using portal imaging.Med Phys. 2003; 30: 2376-2389
- 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT.Radiother Oncol. 2008; 86: 35-42
- A feasible method for clinical delivery verification and dose reconstruction in tomotherapy.Med Phys. 2001; 28: 528-542
- On the accuracy and effectiveness of dose reconstruction for tomotherapy.Phys Med Biol. 2001; 46: 943-966
- A technique for the quantitative evaluation of dose distributions.Med Phys. 1998; 25: 656-661
- Evaluation of the gamma dose distribution comparison method.Med Phys. 2003; 30: 2455-2464
- A revision of the gamma-evaluation concept for the comparison of dose distributions.Phys Med Biol. 2003; 48: 3543-3553
- On dose distribution comparison.Phys Med Biol. 2006; 51: 759-776
- Detection of IMRT delivery errors using a quantitative 2D dosimetric verification system.Med Phys. 2005; 32: 153-162
- Quantitative analysis of patient-specific dosimetric IMRT verification.Phys Med Biol. 2005; 50: 103-119
- An inter-centre quality assurance network for IMRT verification: results of the ESTRO QUASIMODO project.Radiother Oncol. 2005; 76: 340-353
- Interpretation and evaluation of the gamma index and the gamma index angle for the verification of IMRT hybrid plans.Phys Med Biol. 2005; 50: 399-411
- Gamma histograms for radiotherapy plan evaluation.Radiother Oncol. 2006; 79: 224-230
- Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience.Radiother Oncol. 2006; 81: 168-175
- A fast algorithm for gamma evaluation in 3D.Med Phys. 2007; 34: 1647-1654
- Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device.Med Phys. 1999; 26: 2373-2378
- The delivery of intensity modulated radiotherapy to the breast using multiple static fields.Radiother Oncol. 2000; 57: 79-89
- Randomised trial of standard 2D radiotherapy (RT) versus intensity modulated radiotherapy (IMRT) in patients prescribed breast radiotherapy.Radiother Oncol. 2007; 82: 254-264
Article info
Publication history
Published online: August 15, 2008
Accepted:
July 12,
2008
Received in revised form:
July 9,
2008
Received:
April 9,
2008
Identification
Copyright
© 2008 Elsevier Ireland Ltd. Published by Elsevier Inc. All rights reserved.