Radiotherapy & Oncology
Volume 96, Issue 1 , Pages 48-54, July 2010

A study on the dosimetric accuracy of treatment planning for stereotactic body radiation therapy of lung cancer using average and maximum intensity projection images

Department Radiation Oncology, The University of Texas Southwestern Medical Center, Dallas, TX, USA

Received 4 August 2009; received in revised form 5 April 2010; accepted 5 April 2010. published online 30 April 2010.

Abstract 

Purpose

To assess the accuracy of current stereotactic body radiation therapy (SBRT) lung treatment planning methodologies on irregular breathing patterns, we have performed a systematic dosimetric evaluation in phantoms by utilizing maximum intensity projection (MIP) and average (AVG) images generated from four dimensional computed tomography (4DCT).

Methods

A custom built programmable lung phantom was used to simulate tumor motions due to various breathing patterns of patients. 4DCT scans were obtained in helical mode, and reconstructed AVG and MIP datasets were imported into the Pinnacle 8.0h treatment planning system. SBRT plans were generated and executed, and delivered doses were measured by radiochromic film for analysis.

Results

For targets moving regularly or irregularly within a small range (7.0±1.8mm, n=6), we observed good agreement between the measured and computed dose distributions. However, for targets moving irregularly with a larger range (20.8±2.6mm, n=4), the measured isodose lines were found to be shifted relative to the planned distribution, resulting in an under-dosing (over 10%) in a portion of the PTV. We further observed that the discrepancy between planned and measured dose distribution is due to the inaccurate representation of irregular target motion in the MIP images generated from 4DCT.

Conclusions

Caution should be used when planning from 4DCT images in the presence of large and irregular target motion. The inaccuracy inherent in 4DCT MIP and AVG images can be mitigated through the application of methodologies to reduce respiratory motion, such as abdominal compression, and through the use of volumetric image guidance (e.g., cone beam CT – CBCT) to assure precise targeting with minimal shifts.

Keywords: 4DCT, MIP, Stereotactic body radiotherapy, Dosimetry, Lung cancer

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0167-8140(10)00229-X

doi:10.1016/j.radonc.2010.04.003

Radiotherapy & Oncology
Volume 96, Issue 1 , Pages 48-54, July 2010