Journal Home
Search for

Volume 90, Issue 1, Pages 43-47 (January 2009)


View previous. 9 of 30 View next.

Computed tomography for excision cavity localization and 3D-treatment planning in partial breast irradiation with high-dose-rate interstitial brachytherapy

Cynthia AristeiaCorresponding Author Informationemail address, Roberto Tarduccib, Isabella Palumboa, Antonella Cavallia, Francesca Corazzia, Antonio Rullic, Carlo Raymondib, Paolo Latinia

Received 4 October 2007; accepted 17 October 2007. published online 23 November 2007.

Abstract 

Background and purpose

When high-dose-rate brachytherapy is used for partial breast irradiation (PBI) precise pre-implant definition of planning target volume (PTV) and implant geometry is required. After implantation, accurate PTV localization, catheter reconstruction and optimization of dose distribution are needed for good PTV coverage and dose conformity. We applied image-guidance using computed tomography (CT) for pre-implant PTV definition and post-implant dosimetry.

Materials and methods

In 54 patients implant geometry was designed by external beam virtual simulation. A template was placed over dummy beam digitally reconstructed radiographs displaying PTV. Needle entrance and exit points were defined and marked on the patient’s skin to serve as landmarks during implantation. After implantation, in 46/54 patients PTV was defined, catheters were reconstructed and active lengths in the catheters were specified using CT-based-3D planning system. Dosimetry was performed with a Plato-Nucletron treatment planning system.

Results

Post-implantation CT visualized precise catheter placement with respect to the PTV in all patients. CT-based treatment planning provided good coverage of PTV and homogeneous dose distribution.

Conclusions

In post-operative PBI with high-dose-rate brachytherapy CT-based pre-implant definition of implant geometry ensures adequate PTV coverage. After implantation, CT-based 3D-treatment planning software ensures exact PTV localization and catheter reconstruction, and dose distribution optimization.

Article Outline

Abstract

Materials and methods

PTV and implant geometry definition

Before implantation

Implantation day

CT-based 3D-treatment planning

Dosimetry

Therapy

Results

Discussion

Acknowledgment

References

Copyright

Partial breast irradiation (PBI) is more and more often used to treat early breast cancer in patients at low risk of relapse even though it is still considered experimental. When patients receive PBI with high-dose-rate brachytherapy, catheter implantation after lumpectomy requires accurate pre-implant localization of target volume, i.e. the excision cavity plus a 1–2cm margin of surrounding breast tissue [2], and a careful definition of implant geometry in accordance with target volume dimension and shape.

Before implantation pre-operative mammography alone was originally used to localize excision cavity. It is not regarded as sufficient at present, because compression modifies tumour localization. Other modalities have been evaluated. When lumpectomy cavity margins are marked with surgical clips [13], fluoroscopic visualization aids catheter placement [1], [5], [16], [22]. When clips are not positioned at the surgical cavity boundaries [25], ultrasound is a viable option [7], [11], [20] but its accuracy is extremely variable in defining cavity localization and dimensions [3], [7], [9], [17], [18]. Furthermore, the longer the time between surgery and treatment planning, the more ultrasound underevaluates surgical bed dimensions [17]. Das et al. presented an elegant technique for localizing target volume and obtaining adequate coverage with the patient in a prone position on a stereotactic table [6]. The seroma is visualized by ultrasound and its content aspirated. A non-ionic contrast and air are injected to get an air–fluid cavity that is well visualized on the subsequent pre-implant digital mammography. The contrast-enhanced cavity is then centred on a template. The target volume is demarcated and a series of needle positions determined to allow adequate PTV coverage. Despite its advantages, the procedure does not seem to be applied worldwide, mainly due to a lack of necessary instrumentation.

Today, computed tomography (CT) is the best technique for visualizing the surgical cavity [5], [6], [14], [21]. After implantation, CT is essential for treatment planning, as CT-based 3D-treatment planning greatly improves target volume delineation and dosimetry [5], [6], [14].

In this study we describe the CT-based techniques we developed to localize the target volume and design the implant geometry in 54 patients with an open cavity implanted post-operatively. Furthermore, we provide the dosimetric results we obtained when CT-based 3D-treatment planning software was used to delineate the PTV and compute the treatment plan.

Materials and methods 

return to Article Outline

PTV and implant geometry definition 

Before implantation 

For target volume localization the breast is scanned by CT (5mm slice thickness and step in the first 17 patients, 2.5mm in the other 37 because 2.5mm slice thickness and step more accurately define upper and lower cavity edges). The patient is placed in the supine position, with the ipsilateral upper arm extended to the trunk and the forearm bent so the palm of the hand rests upon the head. The same arm position must be used during implantation, post-implant CT and therapy. To ensure position reproducibility 2 points are tattooed at the mediosternal and medium axillary lines. Images are transferred to a treatment planning system (TPS) for external beam radiotherapy (RT) (Pinnacle3 – Philips). The excision cavity is outlined on each slice and, using the TPS volume expansion feature, expanded by 1–2cm to provide the planning target volume (PTV), which is then modified to ensure at least 5mm distance from skin and chest wall.

Expansion varies with breast dimension, lumpectomy cavity location, size and shape. The smaller the breast, the less expansion there must be to avoid overdose to the skin and/or the chest wall and healthy breast tissue. To prevent damage when the cavity is large or close to the chest wall or the skin, expansion might need to be under 1.5–2cm. Depending on the location of irregularly shaped cavities in relation to the skin or chest wall, expansion will probably have to vary at different points. Furthermore, cavity expansion also depends on histological findings as the tumour is not always in the centre of the surgical explant [4]. Consequently, the larger the tumour-free resection margin, the smaller the PTV safety margin. For example, if histological findings report a 3mm margin on one side and a 10mm margin on the other, expansion should be greater on the 3mm side.

Once the PTV has been obtained, a virtual simulation of the treatment plan is carried out using external beam RT software to provide two medial and lateral opposed half beams, the dorsal planes of which become the implant dorsal plane. Medial and lateral digitally reconstructed radiographs (DRRs) of the beams, displaying the PTV, are printed on a scale of 1:1 (Figs. 1A and B). A template is placed over the two DRRs to determine implant needle entrance and exit points.


View full-size image.

Fig. 1. Digitally reconstructed radiographs in which planning target volume (PTV) is displayed. Needle entrance and exit points are determined by placing a template over the PTV. (A) Medial: red=reference point; blue=beam central point.(B) Lateral: green=reference point; orange=beam central point.


Implantation day 

The patient is taken to the simulator. Using the two tattoos as landmarks, the central points of the dummy beams are drawn on the patient’s skin. The dummy beam dorsal planes are then projected and drawn on the skin; needle entrance and exit points are marked (Fig. 2) by placing the template on the patient’s skin.


View full-size image.

Fig. 2. Dummy beam dorsal planes are projected and drawn on the patient’s skin. Entrance and exit points of the superior and inferior needles, previously defined by superimposing a template on the digitally reconstructed radiographs, are marked by placing the template on the patient’s skin.


During implantation, superior and inferior needles on the dorsal plane are inserted by hand, using the markers on the skin as guides. The other needles are inserted using a triangular plastic template (16mm catheter separation; approximately 14mm interplane separation) and a metal template holder to provide accurate needle placement and precise implant geometry. After needle placement, template and holder are removed and the needles are replaced with flexible catheters equipped with buttons blocking both ends.

CT-based 3D-treatment planning 

The patient is taken to the CT room. After trimming and numbering the catheters, X-ray marker wires are inserted, as in standard procedures. The patient undergoes a chest CT scan (5mm slice thickness and step in the first 17 patients, 2.5mm in the others). The tattooed medial and lateral reference points are visualized using radiopaque markers. Images are transferred to a TPS (PLATO – Nucletron). Using a CT-based 3D software (PLATO Brachytherapy Insight) the catheters are identified on three orthogonal planes reconstructed from CT data. The source dwell positions (10mm apart) are localized along the catheters using the X-ray marker wires. Otherwise, the CT slices are scrolled down to visualize catheter buttons. The surgical cavity is outlined on each slice and then expanded, as described before, to get the PTV.

Dosimetry 

Dwell positions are activated in each catheter such that active lengths exceed PTV borders by 2.5 or 5mm. When the PTV border is too close to the skin, active lengths are reduced to spare the skin and reduce the risk of late side effects. Basal points are defined in multiple planes according to active source positions. Optimization on dose points and geometry was performed [19].

The dose is prescribed to the 85% of the mean dose in basal points in the whole volume. Volumes receiving 50%, 100%, 150% and 200% of the prescribed dose (V50, V100, V150 and V200) and conformity index (CI=VR1/TV, where VR1=reference isodose volume and TV=target volume) [8] are calculated by a dose–volume histogram. The dose homogeneity index (DHI) is recognized as the most suitable quality index, and by definition, DHI=1V150/V100 [26]. The percentage dose that covers 90% of the PTV (D90) is also calculated to ensure that no underdosed region exists in the target volume. The percentage of breast tissue, minus the lumpectomy cavity, receiving 100% of prescribed dose is calculated by a dose–volume histogram.

Therapy 

The treatment schedule is 4Gy twice a day for a total dose of 32Gy with a minimum interval of 6h between the 2 daily fractions. Therapy is delivered using a microSelectron HDR 192Ir remote afterloading system (Nucletron).

Results 

return to Article Outline

This approach for localizing the PTV and designing implant geometry was evaluated in 54 patients. A two-plane implant was constructed in 11 cases and a three-plane in 43; median number of implanted catheters was 12 (range 7–15). In all patients post-implantation CT images visualized precise catheter placement with respect to the lumpectomy cavity, so no adjustments were required and the virtual implant geometry was confirmed as accurate.

In the first 8 patients, catheter reconstruction was based on two orthogonal films because the CT-based 3D-treatment planning software was not yet available at our Institution. It was used from September 2004 in the last 46 patients and provided more precise and easier catheter reconstruction, a shorter catheter reconstruction time and more accurate measurements of active and inactive source lengths. Furthermore, the CT-based 3D software visualized dose distribution in the PTV and in organs at risk and calculated the dose–volume histograms. A two-plane implant was constructed in 10 cases and a three-plane implant in 36; the median number of implanted catheters was 12 (range 7–15). Median V100value was 106.0cc (range 43.6–187; mean 106.2; standard deviation – SD – 35.1). Median DHI was 0.75 (range 0.68–0.81; mean 0.76; SD 0.03). Frequency distribution of DHI is reported in Table 1. The mean dose covering 90% of the PTV was 3.86±0.41Gy (96.4%±10.17% of prescribed dose; range 3.08–4.40; median 4.07Gy). Median CI was 1.97 (range 1.50–2.62, mean 1.99, SD 0.46) (Table 2). A mean of 12.39% of healthy breast tissue received 100% of the prescribed dose (range 8.95–15.8, SD 2.58, median 12.75%). Dose distribution around the PTV in a patient with a three-plane implant is shown in Fig. 3.

Table 1.

Dose homogeneity index (DHI): frequency distribution in 46 patients in whom a CT-based-3D planning system was used

DHI
Number of patients
0.68–0.706
0.71–0.735
0.74–0.7619
0.77–0.7914
0.80–0.822
Table 2.

Dosimetric results in 46 patients in whom a CT-based-3D planning system was used

Parameter
Median
Mean
Range
SD
PTV volume (cc)57.957.620.6–106.021.7
V50 (cc)276.5280.0142.0–455.581.6
V100 (cc)106.0106.243.6–18735.1
V150 (cc)27.225.710.6–40.76.5
V200 (cc)12.011.95.3–27.23.7
DHI0.750.760.68–0.810.03
D90 (Gy)4.073.863.08–4.400.41
CI1.971.991.50–2.620.46

SD, standard deviation; PTV, planning target volume; DHI, dose homogeneity index; CI, conformity index.


View full-size image.

Fig. 3. Relative dose distribution of a three-plane implant with PTV on an axial CT slice.


Discussion 

return to Article Outline

When high-dose-rate brachytherapy is used for PBI precise pre-implant localization of the target volume and definition of implant geometry are required [2], [14], [21]. The pre-implant CT-based procedure we developed is accurate because post-implant CT images visualized correct catheter positions with respect to the PTV in all patients so the implant required no adjustments.

CT scanning provides the best level of 3D accuracy in visualizing the surgical cavity [2], [13], [21] for up to 6–8 weeks after surgery [21], i.e. the time-span within which PBI should be delivered. Different CT-image based techniques have been developed for designing a virtual implant geometry before implantation. Perera et al. [15] and Vicini et al. [20] used radiopaque angiocatheters as landmarks for needle entrance and exit points on the breast in a virtual brachytherapy technique. They achieved excellent agreement between pre- and post-implant geometry and no implant adjustment was required in any patient. A more complex, extremely precise technique, described by Cuttino et al. [5], avails of a single section in the CT-simulator suite. After a breast CT scan, the target volume is delineated and a 3D-treatment planning is generated. Catheters are positioned freehand under CT guidance. The entire procedure takes less than 2.5h but cannot be widely adopted because anaesthesia support and a CT simulator are not always available.

Our technique defines implant geometry and catheter placement precisely with respect to the PTV using only two tattoos as landmarks at the mediosternal and medium axillary lines rather than a set of angiocatheters which have to be left in place until the virtual geometry implant has been designed [15], [20]. The tattoos do not cause any discomfort and also serve to reproduce the patient set-up during implantation and subsequent treatment sessions, thus preventing changes in breast position and catheter misalignment. The virtual implant is constructed with external beam TPS and field gantry and collimator angles adjust the implant geometry to the patient’s anatomy. The PTV is localized in the digitally reconstructed radiographs of virtual medial and lateral fields. Guided by the tattoes, the dummy beam central points, the implant dorsal plane and needle entrance and exit points are marked on the patient’s skin.

Like other techniques [5], [15], [20] our simple user-friendly method provides similar results in terms of correct catheter placement with respect to the PTV. As breast interstitial brachytherapy is an operator-dependent technique requiring long training, it is not widely available. CT-based virtual pre-implant geometry construction, such as we describe, renders the procedure less operator dependent and more user-friendly even for radiation oncologists without extensive training in breast brachytherapy.

Treatment quality is guaranteed by precise post-implantation PTV localization, which leads to precise catheter reconstruction, exact inactive and active source length measurement and optimization of dwell weights and positions. All this can be achieved only with image-based 3D information [2], [5], [6], [12], [14].

Treatment planning with CT-based 3D software can provide homogeneous dose distribution in and around the PTV. In our series median D90 values of 4.07Gy show we had achieved good PTV coverage in line with other reports. In Das et al. a median of 96% of target volume received 100% of prescribed dose [6], while Cuttino et al. reported that a mean of 95% PTV (defined as the lumpectomy cavity plus 2cm) received >90% of prescribed dose [5].

A template-guided implant, even with three planes, may not achieve good PTV coverage due to template size [20]. The excellent results obtained by Das et al. who used a template but inserted needles freehand in regions not covered by the template are linked to implantation being performed under cavity visualization [6]. In our series PTV coverage was good. In fact we did not use high-dose-rate brachytherapy for PBI in patients with a large PTV because when we placed the template over the PTV on DRRs we found it did not provide good geometric coverage. Consequently, our mean V100 is smaller than those reported by others (106.2cc vs 211cc achieved by Cuttino et al. and vs 226cc by Das et al.) [5], [6]. We must admit the PTV/V100 ratio is higher than it should have been. Since we used only one type of template to cover PTV we had to irradiate a larger volume of healthy tissue. However, even though the CI was almost double the optimal, it still fell within the range in which treatment is considered to comply with the treatment plan [8]. The disadvantages of the single template will be overcome by using a set with different interplane separation.

DHI is an index of local hot spots in the implant which have been related to fibrosis and fat necrosis [23]. To ensure individual implant quality and homogeneous dose distribution DHI has to be ⩾0.75 and the median value was, in fact, 0.75 in our series (Table 1 for distribution).

Using CT-based 3D software one can visualize the dose distribution in the PTV and the organs at risk on CT slices and calculate the dose–volume histograms. Caution was exercised to prevent unnecessary irradiation of healthy tisssue. Dwell times were modified if too high dose was delivered to the skin so as to prevent teleangiectasis. Dose to the healthy breast tissue was in line with other report (12.39% vs 10% achieved by Weed et al.) [24]. Unlike external beam RT, administered heart and lung doses are far below the limiting value, as demonstrated in the literature [24].

In conclusion, in our series we achieved good coverage of the target volume and homogeneous dose distribution, as demonstrated by the D90 and DHI values. This is a consequence of correct catheter positioning in relation to PTV. In fact, if we had not exactly positioned the catheters, we would probably have been forced to modify dwell times at some dwell position to cover the target volume at the expense of dose homogeneity. We would have had more regions in the target volume receiving over 150% of the planned dose, as reported by Kestin et al. and Major et al. [10], [12]. This would have increased the probability of side effects which, although not life-threatening, impact upon cosmetic results and patient’s quality of life.

Finally, one type of template (with 14mm interplane separation in the present instance) clearly emerges as sub-optimal for all implants. To ensure optimal dose distribution on the PTV and to avoid irradiating healthy tissue unnecessarily, a set of different templates appears needed. Indeed in an ongoing investigation, we have adopted a case-by-case approach and now decide the best strategy for each patient: template with selected interplane separation, freehand, or a combination of template and freehand.

Acknowledgement 

return to Article Outline

The authors would like to thank Dr. Geraldine A. Boyd for translating and editing this paper.

References 

return to Article Outline

[1]. [1]Arthur DW, Koo D, Zwicker RD, et al. Partial breast brachytherapy after lumpectomy: low-dose-rate and high-dose-rate experience. Int J Radiat Oncol Biol Phys. 2003;56:681–689. Abstract | Full Text | Full-Text PDF (89 KB) | CrossRef

[2]. [2]Arthur DW, Vicini FA, Kuske RR, Wazer DE, Nag S. Accelerated partial breast irradiation: an updated report from the American Brachytherapy Society. Brachytherapy. 2003;2:124–130. Abstract | Full Text | Full-Text PDF (113 KB) | CrossRef

[3]. [3]Birdwell RL, Ikeda DM, Torrey MJ, Goffinet DR. Sonographic tailoring of electron beam boost site after lumpectomy and radiation therapy for breast cancer. Am J Roentgenol. 1997;168:39–40.

[4]. [4]Christiaens MR, Cataliotti L, Fentiman I, et al. Comparison of the surgical procedures for breast conserving treatment of early breast cancer in seven EORTC centres. Eur J Cancer. 1996;32A:1866–1875. MEDLINE

[5]. [5]Cuttino LW, Todor D, Arthur DW. CT-guided multi-catheter insertion technique for partial breast brachytherapy: reliable target coverage and dose homogeneity. Brachytherapy. 2005;4:10–17. Abstract | Full Text | Full-Text PDF (336 KB) | CrossRef

[6]. [6]Das RK, Patel R, Shah H, Odau H, Kuske RR. 3D CT-based high-dose-rate breast brachytherapy implants: treatment planning and quality assurance. Int J Radiat Oncol Biol Phys. 2004;59:1224–1228. Abstract | Full Text | Full-Text PDF (187 KB) | CrossRef

[7]. [7]DeBiose DA, Horwitz EM, Martinez AA, et al. The use of ultrasonography in the localization of the lumpectomy cavity for interstitial brachytherapy of the breast. Int J Radiat Oncol Biol Phys. 1997;38:755–759. Full-Text PDF (1159 KB) | CrossRef

[8]. [8]Feuvret L, Noël G, Mazeron J-J, Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys. 2006;64:333–342. Abstract | Full Text | Full-Text PDF (295 KB) | CrossRef

[9]. [9]Gilligan D, Hendry JA, Yarnold JR. The use of ultrasound to measure breast thickness to select electron energies for breast boost radiotherapy. Radiother Oncol. 1994;32:265–267. MEDLINE | CrossRef

[10]. [10]Kestin LL, Jaffray DA, Edmundson GK, et al. Improving the dosimetric coverage of interstitial high-dose-rate breast implants. Int J Radiat Oncol Biol Phys. 2000;46:35–43. Abstract | Full Text | Full-Text PDF (721 KB) | CrossRef

[11]. [11]King TA, Bolton JS, Kuske RR, Fuhrman GM, Scroggins TG, Jiang XZ. Long-term results of wide-field brachytherapy as the sole method of radiation therapy after segmental mastectomy for Tis,1,2 breast cancer. Am J Surg. 2000;180:299–304. Abstract | Full Text | Full-Text PDF (171 KB) | CrossRef

[12]. [12]Major T, Niehoff P, Kovács G, Fodor J, Polgár C. Dosimetric comparison between high dose rate interstitial and MammoSite™ balloon brachytherapy for breast cancer. Radiother Oncol. 2006;79:321–328. Abstract | Full Text | Full-Text PDF (323 KB) | CrossRef

[13]. [13]Nag S, Kuske RR, Vicini FA, Arthur DW, Zwicker RD. Brachytherapy in the treatment of breast cancer. Oncology. 2001;15:195–202. MEDLINE

[14]. [14]Patel RR, Das RK. Image-guided breast brachytherapy: an alternative to whole-breast radiotherapy. Lancet Oncol. 2006;7:407–415. Abstract | Full Text | Full-Text PDF (3385 KB) | CrossRef

[15]. [15]Perera F, Chisela F, Engel J, Venkatesan V. Method of localization and implantation of the lumpectomy site for high dose rate brachytherapy after conservative surgery for T1 and T2 breast cancer. Int J Radiat Oncol Biol Phys. 1995;31:959–965. Abstract | Full-Text PDF (639 KB) | CrossRef

[16]. [16]Polgár C, Major T, Fodor J, et al. High-dose-rate brachytherapy alone versus whole breast radiotherapy with or without tumor bed boost after breast-conserving surgery: seven-year results of a comparative study. Int J Radiat Oncol Biol Phys. 2004;60:1173–1181. Abstract | Full Text | Full-Text PDF (135 KB) | CrossRef

[17]. [17]Rabinovitch R, Finlayson C, Pan Z, et al. Radiographic evaluation of surgical clips is better than ultrasound for defining the lumpectomy cavity in breast boost treatment planning: a prospective clinical study. Int J Radiat Oncol Biol Phys. 2000;47:313–317. Abstract | Full Text | Full-Text PDF (137 KB) | CrossRef

[18]. [18]Ringash J, Whelan T, Elliott E, et al. Accuracy of ultrasound in localization of breast boost field. Radiother Oncol. 2004;72:61–66. Abstract | Full Text | Full-Text PDF (179 KB) | CrossRef

[19]. [19]van der Laarse R, Prins TPE. Introduction to HDR brachytherapy optimisation. In:  Mould RF,  Battermann JJ,  Martinez AA,  Speiser BL editor. Brachytherapy from radium to optimization. Veenendaal: Nucletron International B.V; 1994;p. 331–351.

[20]. [20]Vicini FA, Jaffray DA, Horwitz EM, et al. Implementation of 3D-virtual brachytherapy in the management of breast cancer: a description of a new method of interstitial brachytherapy. Int J Radiat Oncol Biol Phys. 1998;40:629–635. Abstract | Full-Text PDF (5507 KB) | CrossRef

[21]. [21]Wallner P, Arthur D, Bartelink H, et al. Workshop on partial breast irradiation: state of the art and the science, Bethesda, MD, December 8–10, 2002. J Natl Cancer Inst. 2004;96:175–184. CrossRef

[22]. [22]Wazer DE, Berle L, Graham R, et al. Preliminary results of a phase I/II study of HDR brachytherapy alone for T1/T2 breast cancer. Int J Radiat Oncol Biol Phys. 2002;53:889–897. Abstract | Full Text | Full-Text PDF (95 KB) | CrossRef

[23]. [23]Wazer DE, Lowther D, Boyle T, et al. Clinically evident fat necrosis in women treated with high-dose-rate brachytherapy alone for early-stage breast cancer. Int J Radiat Oncol Biol Phys. 2001;50:107–111. Abstract | Full Text | Full-Text PDF (195 KB) | CrossRef

[24]. [24]Weed DW, Edmundson GK, Vicini FA, Chen PY, Martinez AA. Accelerated partial breast irradiation: a dosimetric comparison of three different techniques. Brachytherapy. 2005;4:121–129. Abstract | Full Text | Full-Text PDF (780 KB) | CrossRef

[25]. [25]Wong EK, Truong PT, Kader HA, et al. Consistency in seroma contouring for partial breast radiotherapy: impact of guidelines. Int J Radiat Oncol Biol Phys. 2006;66:372–376. Abstract | Full Text | Full-Text PDF (143 KB) | CrossRef

[26]. [26]Wu A, Ulin K, Sternick ES. A dose homogeneity index for evaluating 192Ir interstitial breast implants. Med Phys. 1988;15:104–107. MEDLINE | CrossRef

a Department of Radiation Oncology, University of Perugia, Italy

b Medical Physics Unit, Perugia General Hospital, Italy

c Department of Surgery, University of Perugia, Perugia, Italy

Corresponding Author InformationCorresponding author. Department of Radiation Oncology, University of Perugia, Policlinico Monteluce, Via Brunamonti, 06122 Perugia, Italy.

PII: S0167-8140(07)00533-6

doi:10.1016/j.radonc.2007.10.029


View previous. 9 of 30 View next.